Feature-Based Registration of Retinal Images

Eli Peli, Reed A. Augliere, George T. Timberlake

IEEE Transactions on Medical Imaging · 1987 · 79 citations · 18 references

Concepts

Abstract

Registration of retinal images taken at different times frequently is required to measure changes caused by disease or to document retinal location of visual stimuli. Cross-correlation has been used previously for such registration, but it is computationally intensive. We have modified a faster algorithm, sequential similarity detection (SSD), to use only the portion of the template that contains retinal vessels. When compared to standard SSD and cross-correlation, this modification improves the reliability of detection for a variety of retinal imaging modalities. The improved reliability enables implementation of a two-stage registration strategy that further decreases the amount of computation and increases the speed of registration.

References

18