Physica Medica · 2020 · 201 citations · 15 references
EngineeringMaterials EngineeringMaterials ScienceNuclear MaterialsRadiation ShieldingSpectroscopyGamma-ray ShieldingGamma-ray SpectroscopyMedical Radiation ShieldingRadiation SpectroscopyShielding MaterialsGamma Radiation ShieldingGamma Shielding MaterialsNuclear Medicine ShieldingRadiation Attenuation CoefficientsRadiation Shielding MaterialsTungsten ShieldingTungsten-based MaterialsTungsten-based Shielding Materials
The study evaluates tungsten carbide as a lead‑free gamma‑ray shielding material for nuclear medicine by assessing its attenuation properties. The authors fabricated tungsten carbide and lead discs of 0.1, 0.5, and 1.0 cm thickness, irradiated them with 123 I, 133 Ba, 152 Eu, and 137 Cs gamma sources (0.160–0.779 MeV), and compared experimental attenuation coefficients, half‑value layers, and mean free paths to NIST theoretical values. Tungsten carbide, 99.9 % pure with 40–50 µm APS, shows experimental mass attenuation coefficients within 0.19–5.11 % of theoretical values and consistently lower half‑value layers and mean free paths than lead across all tested energies.
PurposeThe main objective of this study was to evaluate the efficacy of tungsten carbide as new lead-free radiation shielding material in nuclear medicine by evaluating the attenuation properties.Materials and methodsThe elemental composition of tungsten carbide was analysed using Field-Emission Scanning Electron Microscopy (FESEM) with energy dispersive X-ray (EDX). The purity of tungsten carbide was 99.9%, APS: 40–50 µm. Three discs of tungsten carbide was fabricated with thickness of 0.1 cm, 0.5 cm and 1.0 cm. Three lead discs with similar thickness were used to compare the attenuation properties with tungsten carbide discs. Energy calibration of gamma spectroscopy was performed by using 123I, 133Ba, 152Eu, and 137Cs. Gamma radiation from these sources were irradiated on both materials at energies ranging from 0.160 MeV to 0.779 MeV. The experimental attenuation coefficients of lead and tungsten carbide were compared with theoretical attenuation coefficients of both materials from NIST database. The half value layer and mean free path of both materials were also evaluated in this study.ResultsThis study found that the peaks obtained from gamma spectroscopy have linear relationship with all energies used in this study. The relative differences between the measured and theoretical mass attenuation coefficients are within 0.19–5.11% for both materials. Tungsten carbide has low half value layer and mean free path compared to lead for all thickness at different energies.ConclusionThis study shows that tungsten carbide has high potential to replace lead as new lead-free radiation shielding material in nuclear medicine.
15
The Essential Physics of Medical Imaging
Jerrold T. Bushberg, J. Anthony Seibert, Edwin M. Leidholdt et al. · Medical Physics · 2003 · 2K citations · Full text
Natural Sciences, Physics, Radiology +14