Antibody Modified Gold Electrode as an Impedimetric Biosensor for the Detection of Streptococcus pyogenes

22

Citations

22

References

2020

Year

Abstract

<i>Streptococcus pyogenes</i> is a known cause of a wide spectrum of diseases, from mild and acute to severe invasive infections. This paper concerns the development of a novel impedimetric biosensor for the detection of the mentioned human pathogen. The proposed biosensor is a gold disk electrode modified with commercially available antibodies attached to the surface of the electrode by carbodiimide chemistry. The conducted tests confirmed the specificity of the antibodies used, which was also demonstrated by the results obtained during the detection of <i>S. pyogenes</i> using electrochemical impedance spectroscopy. The developed sensor successfully detected the presence of <i>S. pyogenes</i> in the sample and the detection limit was calculated as 9.3 cfu/mL. The results obtained show a wide linear range for verified concentrations of this pathogen in a sample from 4.2 × 10<sup>2</sup> to 4.2 × 10<sup>6</sup> cfu/mL. Furthermore, the optimal experimentally determined time required to perform pathogen detection in the sample was estimated as 3 min, and the test did not lead to the degradation of the sample.