Pediatric Research · 2020 · 17 citations · 12 references
MedicineDermatologyBiosensorsClinical BiochemistryNeonatologyInfant Skin ConditionsNeonatal JaundiceAccuracyBiological VariationHyperbilirubinemia ScreeningJaundice ScreeningSerum Bilirubin TestingTotal Serum BilirubinTranscutaneous Bilirubin MeasurementTranscutaneous Bilirubinometry
BACKGROUND: Transcutaneous bilirubinometry is a widely used screening method for neonatal hyperbilirubinemia. Deviation of the transcutaneous bilirubin concentration (TcB) from the total serum bilirubin concentration (TSB) is often ascribed to biological variation between patients, but variations between TcB meters may also have a role. This study aims to provide a systematic evaluation of the inter-device reproducibility of TcB meters. MATERIALS AND METHODS: Thirteen commercially available TcB meters (JM-105 and JM-103) were evaluated in vitro on phantoms that optically mimic neonatal skin. The mimicked TcB was varied within the clinical range (0.5-181.3 μmol/L). RESULTS: Absolute differences between TcB meter outcomes increased with the measured TcB, from a difference of 5.0 μmol/L (TcB = 0.5 μmol/L phantom) up to 65.0 μmol/L (TcB = 181.3 μmol/L phantom). CONCLUSION: The inter-device reproducibility of the examined TcB meters is substantial and exceeds the specified accuracy of the device (±25.5 μmol/L), as well as the clinically used TcB safety margins (>50 µmol/L below phototherapy threshold). Healthcare providers should be well aware of this additional uncertainty in the TcB determination, especially when multiple TcB meters are employed in the same clinic. We strongly advise using a single TcB meter per patient to evaluate the TcB over time. IMPACT: Key message: The inter-device reproducibility of TcB meters is substantial and exceeds the clinically used TcB safety margins. What this study adds to existing literature: The inter-device reproducibility of transcutaneous bilirubin (TcB) meters has not been reported in the existing literature. This in vitro study systematically evaluates this inter-device reproducibility. IMPACT: This study aids in a better interpretation of the measured TcB value from a patient and is of particular importance during patient monitoring when using multiple TcB meters within the same clinical department. We strongly advise using a single TcB meter per patient to evaluate the TcB over time.
12
Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation
PEDIATRICS · 2004 · 2.8K citations
Medicine, Child Health, Pediatrics +16
Development of the Epidermis in the Newborn
Nicola Evans, N Rutter · Biology of the Neonate · 2009 · 308 citations
Evaluation of a New Transcutaneous Bilirubinometer
M. Jeffrey Maisels, Enrique M. Ostrea, Suzanne M. Touch et al. · PEDIATRICS · 2004 · 221 citations
Medicine, Pediatrics, Infant Health +12