Publication | Closed Access
Evaluation of a New Transcutaneous Bilirubinometer
220
Citations
33
References
2004
Year
We conclude that TcB measurements using the JM-103 jaundice meter correlate very closely with TSB levels over the range of TSB encountered in this study. Because only 3.3% of our infants had TSB values >15 mg/dL (257 micro mol/L), more data are needed in this range of TSB concentration. The correlation in black infants is not as close as in other groups, but because the tendency in blacks is for the JM-103 to overestimate serum bilirubin levels, dangerous clinical errors are unlikely to occur. The measurement technique is rapid and simple, and it is easy to perform repeated measurements over time, thus reducing the likelihood of error. TcB measurements with the JM-103 jaundice meter should obviate the need for most serum bilirubin levels in newborn infants > or =35 weeks of gestation, although serum bilirubin measurements are still required when treatment with phototherapy or exchange transfusion is being considered.