Health ProfessionsAudiologyEmergency MedicineClinical AudiologyDiagnostic AudiologyHearing AssessmentAir-conduction TestingAudiologic EvaluationHearing ImpairmentHearing LossDiagnostic AudiometryHearing Test InterpretationHearing Threshold MeasurementTest Interpretation In AudiologyThreshold Audiometry
Emergency departments (EDs) are a common location for patients to present with sudden hearing loss (SHL). Unfortunately, high-quality, rapid quantitative measurement of hearing loss is challenging. Herein, we aim to evaluate the accuracy of tablet-based audiometry in patients complaining of SHL. Prospective tablet-based testing was completed in the ED in patients complaining of SHL. Air conduction thresholds (ACTs) obtained via tablet-based audiometry were compared to same-day measurements with a clinical-grade audiometer. Hearing loss (HL) was defined as >20 dB ACT for any frequency. In participant-level analysis, 30+ dB HL in 3 consecutive frequencies was used to define SHL. In the ED, mobile audiogram ACTs were within 5 dB (77%) and 10 dB (89.6%) of those determined by conventional audiometry. The sensitivity and specificity for mobile audiometry to detect 3 or more consecutive thresholds with 30+ dB HL were 100% and 62.5%, respectively. Findings have implications for increasing access to high-quality audiometry.