Biomedical Optics Express · 2020 · 34 citations · 33 references
Natural SciencesBiomedical EngineeringOphthalmologyBiomedical OpticsBiophotonicsOpticsAdaptive OpticsOptical Coherence TomographyOptical ImagingOptical TomographyRetinal ImagingRetinal Pigment EpitheliumAdaptive Optics Optical Coherence TomographyIn Vivo ImagingIn Vivo Retinal ImagingOptical Coherence Tomography InstrumentationOptical Imaging Of CellsOptical Wavefront SensingRetinal LayersWavefront Sensing
Using adaptive optics optical coherence tomography, human photoreceptors and retinal pigment epithelium (RPE) cells are typically visualized on a small field of view of ∼1° to 2°. In addition, volume averaging is required for visualizing the RPE cell mosaic. To increase the imaging area, we introduce a lens based spectral domain AO-OCT system that shows low aberrations within an extended imaging area of 4°×4° while maintaining a high (theoretical) transverse resolution (at >7 mm pupil diameter) in the order of 2 µm. A new concept for wavefront sensing is introduced that uses light mainly originating from the RPE layer and yields images of the RPE cell mosaic in a single volume acquisition. The capability of the instrument for in vivo imaging is demonstrated by visualizing various cell structures within the posterior retinal layers over an extended field of view.
33
Thomas Klein, Wolfgang Wieser, Lukas Reznicek et al. · Biomedical Optics Express · 2013 · 216 citations · Full text