Optics Express · 2005 · 432 citations · 27 references
Natural SciencesBiomedical EngineeringOphthalmologyBiomedical ImagingBiomedical OpticsBiophotonicsAdaptive OpticsOphthalmic ImagingOptical CoherenceOptical Coherence TomographyRetinal AnatomyRetinal ImagingAdaptive Optics Optical Coherence TomographyAxial ResolutionCone Photoreceptor MosaicFourier-domain Optical Coherence TomographyLateral ResolutionShack-hartmann Wavefront SensorsWavefront Aberration CorrectionWavefront Sensing
We have combined Fourier-domain optical coherence tomography (FD-OCT) with a closed-loop adaptive optics (AO) system using a Hartmann-Shack wavefront sensor and a bimorph deformable mirror. The adaptive optics system measures and corrects the wavefront aberration of the human eye for improved lateral resolution (~4 μm) of retinal images, while maintaining the high axial resolution (~6 μm) of stand alone OCT. The AO-OCT instrument enables the three-dimensional (3D) visualization of different retinal structures in vivo with high 3D resolution (4×4×6 μm). Using this system, we have demonstrated the ability to image microscopic blood vessels and the cone photoreceptor mosaic.
27
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Human photoreceptor topography
Christine A. Curcio, Kenneth R. Sloan, Robert E. Kalina et al. · The Journal of Comparative Neurology · 1990 · 2.8K citations
Medicine, Anatomy, Vision Science +15