Detecting trace methane levels with plasma optical emission spectroscopy and supervised machine learning

Jordan Vincent, Hui Wang, Omar Nibouche, Paul Maguire

Plasma Sources Science and Technology · 2020 · 15 citations · 27 references

DOIFull text

Open access

Concepts

Abstract

Abstract Trace methane detection in the parts per million range is reported using a novel detection scheme based on optical emission spectra from low temperature atmospheric pressure microplasmas. These bright low-cost plasma sources were operated under non-equilibrium conditions, producing spectra with a complex and variable sensitivity to trace levels of added gases. A data-driven machine learning approach based on partial least squares discriminant analysis was implemented for CH 4 concentrations up to 100 ppm in He, to provide binary classification of samples above or below a threshold of 2 ppm. With a low-resolution spectrometer and a custom spectral alignment procedure, a prediction accuracy of 98% was achieved, demonstrating the power of machine learning with otherwise prohibitively complex spectral analysis. This work establishes proof of principle for low cost and high-resolution trace gas detection with the potential for field deployment and autonomous remote monitoring.

References

27