Journal of Separation Science · 2020 · 13 citations · 10 references
Fast, simple, accurate, and reproducible reverse phase-high-performance liquid chromatography method with direct ultraviolet measurement of memantine hydrochloride in tablets was developed, without any chemical derivatization pretreatment. Three main problems appear during chromatographic analysis of memantine: detection, achieving appropriate column retention, and limited choice of mobile phase components, as a result of memantine molecular structure. Among more than 35 tested columns, the best retention and peak symmetry yielded two C8 and three C18 columns with different characteristics, at a temperature of 30°C, mobile phase composed of 1%, v/v, acetonitrile and 99%, v/v, of 0.05-0.1% phosphoric acid or 2.5-5 mmol phosphate buffer, at flow rate of 1 mL/min and injection volume of 5 µL. The retention time of memantine was between 2.6 and 4 min. Both mobile phase concepts showed perfect linearity, precision, and accuracy. This is the first successful and reproducible direct reverse phase-high-performance liquid chromatography-ultraviolet quantification method for memantine.
10
Bhavil Narola, Amit Singh, P. Rita Santhakumar et al. · Analytical Chemistry Insights · 2010 · 41 citations · Full text
Ariadne M. Brondi, Jerusa S. Garcia, Marcello G. Trevisan · Journal of Analytical Methods in Chemistry · 2017 · 17 citations · Full text