Publication | Open Access
Collinear setup for delay control in two-color attosecond measurements
12
Citations
22
References
2020
Year
EngineeringOptical TestingMid-infrared Laser TechnologyCompact Experimental SetupOptical PropertiesOptical DiagnosticsInfrared OpticInstrumentationOptical SpectroscopyTwo-photon PhotoionizationLong-time Delay StabilityOptical PumpingPhotonicsPhysicsInfrared SpectroscopyPhotoelectric MeasurementNatural SciencesSpectroscopyApplied PhysicsCollinear Setup
Abstract We present a compact experimental setup for performing attosecond-pump-infrared-probe experiments with long-time delay stability. The robustness of the setup is demonstrated over a two-day acquisition time in two-photon photoionization of argon in the photon-energy range 17−33 eV. The propagation of the input infrared pulse, as driving pulse for the high-order harmonic generation process and for the generation of the sidebands of the main photoelectron peaks, through the main optical components is simulated and discussed. Our setup allows us to perform attosecond experiments with an overall stability of ± 40 as.
| Year | Citations | |
|---|---|---|
Page 1