Influence of Irradiation Temperature on Oxidative and Network Properties of X‐Ray Cross‐Linked Vitamin E Stabilized UHMWPE for Hip Arthroplasty

Marie Anne Mulliez, Christoph Schilling, Thomas M. Grupp

BioMed Research International · 2020 · 15 citations · 48 references

DOIFull text

Open access

Concepts

Abstract

Previous studies have shown that increased cross‐link density, reduced free radicals, and increased antioxidant grafting resulting from electron‐beam irradiation at elevated temperatures improved the wear performance and the oxidative stability of vitamin E blended UHMWPE. The current study explores the impact of elevated irradiation temperature on vitamin E blended UHMWPE using X‐ray. We hypothesize that the effects of temperature would be similar to those observed after electron‐beam irradiation due to the relatively high dose rate of X‐rays. Two X‐ray doses of 80 and 100 kGy and two irradiation temperatures, that is, room temperature and 100°C were considered. The reference was Vitelene®, a vitamin E stabilized polyethylene cross‐linked with 80 kGy by e‐beam at 100°C. Oxidation index and oxidation induction time, as well as cross‐link density, gel fraction, and trans ‐vinylene index, were determined, as the oxidative and network properties are decisive for the long‐term implant performance. Gel fraction and oxidation induction time were significantly improved subsequently to warm irradiation in comparison with the material irradiated at room temperature. In conclusion, X‐ray irradiation at elevated temperatures resulted in an increase of cross‐linking and oxidative resistance of vitamin E stabilized polyethylene comparable to those of e‐beam irradiated UHMWPE.