Biomedical Optics Express · 2020 · 61 citations · 32 references
MedicineBioengineeringNeuroscienceBiomedical ImagingBiomedical Signal ProcessingClinical NeuroscienceBrain ImagingCerebral Blood FlowCerebral PhysiologyHemodynamicsIntracranial PressureIntracranial Pressure MonitoringCerebral Blood Flow AlterationsCerebral Blood Flow QuantificationIncreased Intracranial PressureIntracranial HemodynamicsIntracranial Pressure DynamicsNoninvasive Intracranial Pressure EstimationNoninvasive Intracranial Pressure MonitoringPulsatile Flow
Measuring intracranial pressure (ICP) is necessary for the treatment of severe head injury but measurement systems are highly invasive and introduce risk of infection and complications. We developed a non-invasive alternative for quantifying ICP using measurements of cerebral blood flow (CBF) by diffuse correlation spectroscopy. The recorded cardiac pulsation waveform in CBF undergoes morphological changes in response to ICP changes. We used the pulse shape to train a randomized regression forest to estimate the underlying ICP and demonstrate in five non-human primates that DCS-based estimation can explain over 90% of the variance in invasively measured ICP.
32
Leo Breiman · Machine Learning · 2001 · 126.2K citations · Full text
Scikit-learn: Machine Learning in Python
Fabián Pedregosa, Gaël Varoquaux, Alexandre Gramfort et al. · arXiv (Cornell University) · 2012 · 63.7K citations · Full text
Intracranial Pressure Monitoring: Invasive versus Non-Invasive Methods—A Review
P. H. Raboel, Jir̂í Bártek, Morten Andresen et al. · Critical Care Research and Practice · 2012 · 419 citations · Full text
Medicine, Neurology, Neurosurgery +14