Concepedia

Publication | Open Access

Electrodermal activity patient simulator

24

Citations

37

References

2020

Year

Abstract

Electrodermal activity (EDA) is an electrical property of the human skin, correlated with person's psychological arousal. Nowadays, different types of EDA measuring devices are used in highly versatile fields-from research, health-care and education to entertainment industry. But despite their universal use the quality of their measuring function (their accuracy) is questioned or investigated very seldom. In this paper, we propose a concept of an EDA patient simulator-a device enabling metrological testing of EDA devices by means of a variable resistance. EDA simulator was designed based on a programmable light-controlled resistor with a wide resistance range, capable of simulating skin conductance levels (SCL) and responses (SCR) and was equipped with an artificial hand. The hand included electrically conductive fingers for attachment of EDA device electrodes. A minimal set of tests for evaluating an EDA device was identified, the simulator's functionality discussed and some testing results presented.

References

YearCitations

Page 1