IEEE Sensors Journal · 2020 · 19 citations · 20 references
EngineeringNuclear EngineeringNuclear MedicineRadiation DetectionRadiation DosimetryRadiation MeasurementDetector EfficiencyDetector TechnologyDosimetry InstrumentationGamma-ray DetectionRadiation DetectorsScintillation DetectorsDose RepeatabilityDose-rate EffectsInitial Performance EvaluationRepeatabilityRepeatability AssessmentScintillation DetectionScintillation DosimetrySetup Uncertainty Analysis
The purpose of this study was to design and evaluate the performance of four novel inorganic scintillating detectors (ISDs) on the Small Animal Radiation Research Platform (SARRP). Relative scintillator output, measurement repeatability, setup uncertainty, linearity with dose rate, and signal reproducibility over time were investigated. The Gd <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> S:Tb detector had the highest relative signal output, generating up to 219 times more charge than a previously characterized BCF-60-based plastic scintillating detector (PSD). The Gd <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> S:Tb detector was then used to measure 220 kVp therapy beam profiles of 10 x 10 and 5 x 5 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> fields. Beam profiles using the ZnS-based phosphor were also obtained and compared to investigate the performance of a lower density inorganic scintillator. 10 x 10 and 5 x 5 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> therapy beam profile measurements made with the Gd <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> S:Tb and BCF-60 detectors differed, on average, by 1.1% and 1.9%, respectively. The ZnS:Ag measurements differed, on average, by 2.5% and 6% relative to BCF-60 measurements of the 10 x 10 and 5 x 5 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> beam profiles, respectively. MicroCT imaging of the detector volumes was also performed, revealing poor packing of the ZnS:Ag crystalline phosphor in the deepest region of the cylindrical cavity. The Gd <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> S:Tb detector, in particular, has proven to be a promising candidate for real-time dosimetry of small fields in small animal irradiators, primarily because of the very large signal intensities observed, along with good repeatability, dose rate linearity, reproducibility and agreement with beam profile measurements made with a previously validated detector.
20
Scintillation detectors for x-rays
M. Nikl · Measurement Science and Technology · 2006 · 852 citations
Experimental correction for the inner-filter effect in fluorescence spectra
Mikael Kubista, Robert Sjöback, Svante Eriksson et al. · The Analyst · 1994 · 398 citations
John Wong, Elwood P. Armour, Peter Kazanzides et al. · International Journal of Radiation Oncology*Biology*Physics · 2008 · 361 citations · Full text
Veterinary Medicine, Radiology, Veterinary Diagnostic Imaging +14