Applied Optics · 2020 · 28 citations · 26 references
EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingDigital Image ProcessingImage EnhancementImage FilteringImage QualityImage ReconstructionMedical ImagingArtifact ReductionContrastDigital FilteringFilteringImage ResamplingSignal-to-noise RatioSignal-to-noise Ratio EnhancementSpatial Resampling
Fiber bundles have become widely adopted for use in endoscopy, live-organism imaging, and other imaging applications. An inherent consequence of imaging with these bundles is the introduction of a honeycomb-like artifact that arises from the inter-fiber spacing, which obscures features of objects in the image. This artifact subsequently limits applicability and can make interpretation of the image-based data difficult. This work presents a method to reduce this artifact by on-axis rotation of the fiber bundle. Fiber bundle images were first low-pass and median filtered to improve image quality. Consecutive filtered images with rotated samples were then co-registered and averaged to generate a final, reconstructed image. The results demonstrate removal of the artifacts, in addition to increased signal contrast and signal-to-noise ratio. This approach combines digital filtering and spatial resampling to reconstruct higher-quality images, enhancing the utility of images acquired using fiber bundles.
26
High-speed, miniaturized fluorescence microscopy in freely moving mice
Benjamin A. Flusberg, Axel Nimmerjahn, Eric D. Cocker et al. · Nature Methods · 2008 · 395 citations