Prospective Analysis of the Impact of Commercialized Hybrid Closed-Loop System on Glycemic Control, Glycemic Variability, and Patient-Related Outcomes in Children and Adults: A Focus on Superiority Over Predictive Low-Glucose Suspend Technology

Pilar Isabel Beato-Víbora, Fabiola Gallego-Gamero, Lucía Lázaro-Martín, María del Mar Romero-Pérez, Francisco Javier Arroyo-Díez

Diabetes Technology & Therapeutics · 2019 · 66 citations · 22 references

Abstract

<b><i>Background:</i></b> Automatization of insulin delivery by closed-loop systems represents a major step in type 1 diabetes management. The aim of this study was to analyze the effect of the commercialized hybrid closed-loop system, the MiniMed 670G system, on glycemic control, glycemic variability, and patient satisfaction. <b><i>Methods:</i></b> A prospective study, including type 1 diabetes patients consecutively starting on the 670G system in one adult and two pediatric hospitals, was performed. Baseline and 3-month visits were documented. Two weeks of data from the system were downloaded. Glycemic variability measures were calculated. Adults and adolescents completed a set of questionnaires (Gold and Clarke scores, Hypoglycemia Fear Survey, Diabetes Quality of Life [DQoL], Diabetes Treatment Satisfaction [DTS], Diabetes Distress Scale, Pittsburgh Sleep Quality Index). <b><i>Results:</i></b> Fifty-eight patients were included (age: 28 ± 15 years [7-63], <18 years old: 38% [<i>n</i> = 22], 59% [<i>n</i> = 34] females, previous use of SAP-PLGS [predictive low-glucose suspend]: 60% [<i>n</i> = 35]). HbA1c was reduced from 57 ± 10 to 53 ± 7 mmol/L (7.4% ± 0.9% to 7.0% ± 0.6%) (<i>P</i> < 0.001) and time in range 70-180 mg/dL was increased from 63.0% ± 11.4% to 72.7% ± 8.7% (<i>P</i> < 0.001). In patients with high baseline hypoglycemia risk, time <54 and <70 mg/dL were reduced from 0.9% ± 1.1% to 0.45% ± 0.7% (<i>P</i> = 0.021) and from 3.3% ± 2.8% to 2.1% ± 2.1% (<i>P</i> = 0.019), respectively. Glycemic variability measures improved. Time in auto mode was 85% ± 17%, the number of auto mode exits was 0.6 ± 0.3 per day, and the number of alarms was 8.5 ± 3.7 per day. Fear of hypoglycemia, DQoL, DTS, and diabetes distress improved, while the percentage of patients with poor sleep quality was reduced. The discontinuation rate was 3%. <b><i>Conclusion:</i></b> The commercialized hybrid closed-loop system improves glycemic control and glycemic variability in children and adults, reducing the burden of living with type 1 diabetes.

References

22