Battery‐free implantable insulin micropump operating at transcutaneously radio frequency‐transmittable power

Bingxi Yan, Duo An, Xi Wang, Brock J. DeLong, Asimina Kiourti, Kathleen Dungan, John L. Volakis, Minglin Ma, Liang Guo

Medical Devices & Sensors · 2019 · 24 citations · 14 references

DOIFull text

Open access

Concepts

Abstract

Abstract Implantable insulin pumps that can be wirelessly powered may provide a pain‐free solution to continual treatment of Type‐1 diabetes. However, their design and development have long been prohibited by the lack of low‐voltage actuators operated at a wirelessly deliverable small power. Here we introduce a battery‐free insulin pump actuated by a soft and inherently biocompatible polymeric squeezer. At a low voltage below 1.5 V and a small power below 2 mW, which is deliverable using a miniature wireless harvester, our pump consistently delivers a volumetric output of 10 µl within 1 min at a flow rate of 0.21 µl/s. Moreover, comparative studies based on diabetic mice show well agreed blood glucose responses of pump‐ and syringe‐delivery solutions. These encouraging results pave the way for a lightweight, small‐dimension and battery‐free solution to overcome major disadvantages in current implantable insulin pumps.

References

14