Journal of Radiological Protection · 2019 · 40 citations · 22 references
MedicineOphthalmologyRadiologyDiagnostic RadiologyInterventional RadiologyRadiation DosimetryDosimetry InstrumentationPersonal DosimetryPersonnel DosimetryPhoton DosimetryRadiation Dose MeasurementAngular DependenceBatch UniformityDosimeter Energy DependenceDosimetry In Interventional ProceduresEnergy DependenceEye Lens DosimetryRadiation Dose EstimationX-ray Dosimetry
Monitoring and protecting of occupational eye doses in interventional radiology (IR) are very important matters. DOSIRIS™ is the useful solution to estimate the 3 mm dose-equivalent (Hp(3)), and it can be worn behind lead glasses. And DOSIRIS™, adjustable according to 3 axes, it is ideally placed as close to the eye and in contact with the skin. So, DOSIRIS™ will be suitable eye lens dosimeter. However, the fundamental characteristics of the DOSIRIS™ in the diagnostic x-ray energy domain (including that of IR x-ray systems) remain unclear. Here, we evaluated the performance of the dosimeter in that energy range. As a result, the DOSIRIS™ has good fundamental characteristics (batch uniformity, dose linearity, energy dependence, and angular dependence) in the diagnostic x-ray energy domain. We conclude that the DOSIRIS™ has satisfactory basic performance for occupational eye dosimetry in diagnostic x-ray energy settings (including IR x-ray systems).
22
Occupational eye dose in interventional cardiology procedures
Yoshihiro Haga, Koichi Chida, Yuji Kaga et al. · Scientific Reports · 2017 · 187 citations · Full text
Occupational Dose in Interventional Radiology Procedures
Koichi Chida, Yuji Kaga, Yoshihiro Haga et al. · American Journal of Roentgenology · 2012 · 175 citations