A fast volumetric 4D-MRI with sub-second frame rate for abdominal motion monitoring and characterization in MRI-guided radiotherapy

Jing Yuan, Oi Lei Wong, Yihang Zhou, Kin Yin Chueng, Siu Ki Yu

Quantitative Imaging in Medicine and Surgery · 2019 · 26 citations · 44 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: To propose a fast volumetric 4D-MRI based on 3D pulse sequence acquisition for abdominal motion monitoring and characterization in MRI-guided radiotherapy (MRgRT). METHODS: , were acquired in 89 s for 4D-MRI. This CAIPIRINHA-VIBE 4D-MRI was qualitatively compared with a 2D half-Fourier acquisition single-shot turbo spin-echo (2D-HASTE) based 4D-MRI. The motions of liver dome, kidney and spleen were analyzed using the CAIPIRINHA-VIBE 4D-MRI data. The kidney motion was quantitatively characterized in terms of motion range and the correlations between left and right kidneys. RESULTS: 2.32±1.34 mm, P=0.045) than left kidney in AP, but not in LR and SI directions. The Pearson correlation coefficients r between left and right kidney motion were 0.5063 (P=0.164), 0.6624 (P=0.052) and 0.5752 (P=0.105) in LR, AP and SI correspondingly. The correlation of renal motion in SI and AP was found significant in right kidney (r=0.843, P=0.004) but not in left kidney (r=0.467, P=0.205). CONCLUSIONS: A fast volumetric 4D-MRI was implemented for abdominal motion monitoring in MRgRT. A sub-second volumetric temporal resolution of 0.615 s, covering the entire abdomen, was demonstrated for respiratory motion monitoring and characterization. This technique holds potentials for MRgRT applications.