Magnetic Resonance in Medicine · 2019 · 13 citations · 35 references
EngineeringBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage ReconstructionMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMedical ImagingMedical Imaging InstrumentationRenal Imaging3D Magnetic Resonance ImagingIn Vivo Hyperpolarized ImagingParallel ImagingParallel Imaging ReconstructionParallel Imaging SequencesThree-dimensional Image ReconstructionTissue Inhomogeneity CorrectionVolumetric Data Acquisition
Purpose To investigate auto‐ and pre‐calibration coil profile estimation for parallel imaging reconstruction of hyperpolarized 13 C MRI volumetric data. Methods Parallel imaging reconstruction was studied with 3 different approaches for coil profile estimation: auto‐calibration, phantom calibration, and theoretic calibration. Acquisition was performed with a 3D stack‐of‐spirals sequence with spectral–spatial excitation and Cartesian undersampling. Parallel imaging reconstructions were done with conjugate gradient SENSE and 3D gridding with inhomogeneity correction. The approaches were compared in simulations with different SNR, through phantom experiments, and in an in vivo pig study focused on the kidneys. All imaging was done with a rigid home‐built 12‐channel 13 C receive coil at 3T. Results The phantom calibrated and theoretic approaches resulted in the best structural similarities in simulations and demonstrated higher image quality in the phantom experiments compared to the auto‐calibrated approach. In vivo mapping of pyruvate uptake and lactate conversion improved for accelerated acquisitions because of a better temporal resolution. From a practical and image quality point of view, use of theoretic coil profiles led to improved results compared to the other approaches. Conclusion The success of the theoretic coil profile estimation demonstrates a negligible effect of load on sensitivity profiles at the carbon frequency at 3T. Through theoretic or phantom calibrated parallel imaging, accelerated 3D volumes could be reconstructed with sufficient sensitivity, temporal, and spatial resolution to map the metabolism of kidneys exemplifying abdominal organs. This approach overcomes a critical step in the clinical translation of parallel imaging in hyperpolarized 13 C MR.
35
Image quality assessment: from error visibility to structural similarity
Zhou Wang, Alan C. Bovik, Hamid R. Sheikh et al. · IEEE Transactions on Image Processing · 2004 · 56K citations
SENSE: Sensitivity encoding for fast MRI
Klaas P. Pruessmann, Markus Weiger, Markus B. Scheidegger et al. · Magnetic Resonance in Medicine · 1999 · 6.1K citations
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Mark A. Griswold, Peter M. Jakob, Robin M. Heidemann et al. · Magnetic Resonance in Medicine · 2002 · 5.3K citations · Full text