Photoacoustics · 2019 · 46 citations · 30 references
MedicineBiomedical EngineeringBiomedical ImagingUltrasonographyUltrasoundDiagnostic UltrasonographyPhotoacoustic ImagingTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound InstrumentationLead Zirconate TitanatePhotoacoustic ProbesPhotoacoustic TransducersSignal-to-noise RatioSignal-to-noise Ratio EnhancementTransducersUltrasonic ImagingUltrasonic Transducers
Endoscopic dual-modality photoacoustic (PA) and ultrasound (US) imaging has the capability of providing morphology and molecular information simultaneously. An ultrasonic transducer was applied for detecting PA signals and performing US imaging which determines the sensitivity and performance of a dual-modality PA/US system. In our study, a miniature single element 32-MHz lead magnesium niobate-lead titanate (PMN-PT) epoxy 1–3 composite based ultrasonic transducer was developed. A miniature endoscopic probe based on this transducer has been fabricated. Using the dual modality PA/US system with a PMN-PT/epoxy 1–3 composite based ultrasonic transducer, phantom and in vivo animal studies have been conducted to evaluate the performance. The preliminary results show enhanced bandwidths of the new ultrasonic transducer and improved signal-to-noise ratio of PA and US images of rat colorectal wall compared with PMN-PT and lead zirconate titanate (PZT) composite based ultrasonic transducers.
30
Looking at sound: optoacoustics with all-optical ultrasound detection
Georg Wissmeyer, Miguel A. Pleitez, Amir Rosenthal et al. · Light Science & Applications · 2018 · 369 citations · Full text
Joon-Mo Yang, Konstantin Maslov, Hao-Chung Yang et al. · Optics Letters · 2009 · 224 citations · Full text