Colorimetric detection of hydrogen peroxide and glucose by exploiting the peroxidase-like activity of papain

Yuye Chen, Qingmei Zhong, Yilin Wang, Chunling Yuan, Xiu Qin, Yuanjin Xu

RSC Advances · 2019 · 62 citations · 44 references

DOIFull text

Open access

Concepts

Abstract

in the range of 5.00-90.0 μM, along with a detection limit of 2.10 μM. A facile colorimetric method for glucose detection was also proposed by combining the glucose oxidase (GOx)-catalyzed glucose oxidation and papain-catalyzed TMB oxidation, which exhibited a linear response in the range of 0.05-0.50 mM with a detection limit of 0.025 mM. The method proposed here displayed excellent selectivity, indicating that common coexisting substances (urea, uric acid, ascorbic acid, maltose, lactose and fructose) in urine did not interfere with detection of glucose. More importantly, the suggested method was successfully used to precisely detect the glucose concentration in human urine samples with recoveries over 96.0%.

References