Novel design for a customized, 3D‐printed surgical template for thoracic spinal arthrodesis

Francesco Naddeo, Carlotta Fontana, Alessandro Naddeo, Emilio Cataldo, Nicola Cappetti, Nicola Narciso

International Journal of Medical Robotics and Computer Assisted Surgery · 2019 · 17 citations · 24 references

Concepts

Abstract

BACKGROUND: The integration of computer-aided design/computer-aided manufacturing (CAD/CAM) tools and medicine is rapidly developing for designing medical devices. A novel design for a 3D-printed patient-specific surgical template for thoracic pedicle screw insertion, using a procedure based on reverse engineering, is presented. METHODS: The surgeon chooses the entry point on the vertebra. The optimal insertion direction and the size of the screws are defined via an algorithm on the basis of a patient-specific vertebra CAD model. The template features an innovative shape for a comfortable and univocal placement and a novel disengaging device. RESULTS: Three spinal fusions were performed to test the template. Excellent results were achieved in terms of the accuracy of the screw positioning, reduction in surgery duration, and number of X-rays. CONCLUSIONS: A novel design for a customized, 3D-printed surgical template for thoracic spinal arthrodesis was presented, and improvements in terms of precision, duration, and safety were achieved without changing the standard procedure.

References

24