International Journal of Medical Robotics and Computer Assisted Surgery · 2019 · 17 citations · 24 references
MedicineOrthopedic SurgerySurgeryAdditive ManufacturingOrthopedicsSpine SurgeryDevice DesignThoracic Spine Surgery3D Printed Surgical ModelsManufacturingPlacementSurgical Template DesignSurgical Template FabricationThoracic Pedicle Screw FixationThoracic Spinal Fusion
BACKGROUND: The integration of computer-aided design/computer-aided manufacturing (CAD/CAM) tools and medicine is rapidly developing for designing medical devices. A novel design for a 3D-printed patient-specific surgical template for thoracic pedicle screw insertion, using a procedure based on reverse engineering, is presented. METHODS: The surgeon chooses the entry point on the vertebra. The optimal insertion direction and the size of the screws are defined via an algorithm on the basis of a patient-specific vertebra CAD model. The template features an innovative shape for a comfortable and univocal placement and a novel disengaging device. RESULTS: Three spinal fusions were performed to test the template. Excellent results were achieved in terms of the accuracy of the screw positioning, reduction in surgery duration, and number of X-rays. CONCLUSIONS: A novel design for a customized, 3D-printed surgical template for thoracic spinal arthrodesis was presented, and improvements in terms of precision, duration, and safety were achieved without changing the standard procedure.
24
A novel computer-assisted drill guide template for placement of C2 laminar screws
Sheng Lu, Yong Xu, Yuan Z. Zhang et al. · European Spine Journal · 2009 · 100 citations · Full text