Current Sample Preparation and Analytical Capabilities of the Radiocarbon Laboratory at CologneAMS

Janet Rethemeyer, Merle Gierga, Stefan Heinze, A. Stolz, Anja Wotte, Philipp Wischhöfer, Sonja Berg, JO Melchert

Radiocarbon · 2019 · 60 citations · 26 references

Concepts

Abstract

ABSTRACT This work summarizes the methodical capabilities, improvements, and new developments in the radiocarbon laboratory of the accelerator mass spectrometry (AMS) facility at the University of Cologne, Germany, which was established in 2010. During the past years, the laboratory has specialized in the analysis of small and gaseous samples. We thus, recently installed a second ion source dedicated for radiocarbon ( 14 C) analysis of CO 2 samples at our 6 MV Tandetron AMS from High Voltage Engineering Europe B.V. that is coupled with the gas injection system from Ionplus and an EuroVector EA 3000 elemental analyzer. This work summarizes all pretreatment methods and analytical facilities established in our laboratory during the last years including 14 C analysis of individual organic compounds and of CO 2 trapped on molecular sieves. We also report different blank values including our long-term blank since 2011, which is for normal-sized, solid samples (650–1000 µg C) 0.0012 ± 0.0004 F 14 C (54,305 ± 2581 yr BP, n = 484). The precision obtained for modern samples measured as graphite is 0.5% and for gaseous samples injected with the GIS ≤2%.