Microscopy and Microanalysis · 2019 · 13 citations · 22 references
Natural SciencesPhysicsAtomic PhysicsBiomedical ImagingDiagnostic ImagingCalibrationComputed TomographyComputed Tomography PhysicsImage ReconstructionAccuracyAdaptive ReconstructionReconstruction Error AnalysisSpatial Accuracy AssessmentSpatial Error ModelsTomographic ImagingTomographic QuantificationTomographic Reconstruction
We define a measure for the accuracy of tomographic reconstruction in atom probe tomography, named here the spatial error index. We demonstrate that this index can be used to compare rigorously the spatial accuracy of various different approaches to the calculation of tomographic reconstruction. This is useful, for example, to evaluate the performance of alternate tomographic reconstruction approaches, and ensures that the comparisons are independent of individual data quality or other instrumental parameters. We then introduce a new "adaptive reconstruction" formalism that uses a progression of reconstruction parameters based on a per-atom correction from the cube root of the inverse of the voltage, along with linear correction factors linked to the evaporation sequence. We apply the measure for spatial accuracy to this new reconstruction protocol.
22
A general protocol for the reconstruction of 3D atom probe data
Pierre‐Yves Le Bas, A. Bostel, B. Déconihout et al. · Applied Surface Science · 1995 · 428 citations
Advances in the reconstruction of atom probe tomography data
Baptiste Gault, Daniel Haley, Frédéric De Geuser et al. · Ultramicroscopy · 2010 · 238 citations
Spatial Distribution Maps for Atom Probe Tomography
B. Geiser, Thomas F. Kelly, David J. Larson et al. · Microscopy and Microanalysis · 2007 · 192 citations