Review of Scientific Instruments · 2019 · 11 citations · 10 references
EngineeringBiomedical EngineeringMaterials ScienceBiomedical ImagingRadiation DetectionSurface ScienceElectron Beam PhysicsElectron MicroscopyRadiation DetectorsRadiation Imaging DetectorsSemiconductor DetectorsMicrostructureScanning Electron MicroscopySignal-to-noise RatioSubsurface Structure ImagingSubsurface Structure MappingSurface ReliefSurface Relief MeasurementTomography
A new configuration of semiconductor detectors for backscattered electrons for a scanning electron microscope (SEM) is presented. The result of the optimization was the possibility to extract the information about the spatial relief (3D topology) of the sample and its subsurface structure (3D tomography) in the simplest way. The detector consists of 8 sensors-semiconductor plates, positioned in a certain way. The proposed method was tested on real structures having a surface micro relief or a subsurface volume structure. Experiments and simple calculations show increased effectiveness and a high signal-noise ratio in the proposed method. This is important, particularly for studying the radiation-sensitive biomedical tissue in SEM.
10
Electron backscattering from thin films
Heinz Niedrig · Journal of Applied Physics · 1982 · 223 citations