Spreading a water droplet through filter paper on the metal substrate for surface-enhanced laser-induced breakdown spectroscopy

Xinyan Yang, Rongxing Yi, Xiangyou Li, Zhifeng Cui, Yongfeng Lu, Zhongqi Hao, Jingchun Huang, Zhengxian Zhou, Guanxin Yao, Wanxia Huang

Optics Express · 2018 · 38 citations · 45 references

DOIFull text

Open access

Concepts

Abstract

values were improved from 0.6192~0.9321 to 0.9481~0.9766, the RMSECV values were decreased from 0.53~1.95 μg/mL to 0.33~1.06 μg/mL, and the average relative error (ARE) values were decreased from 8.96~22.31% to 4.28~14.37% with the aid of filter paper. This demonstrated that the use of filter paper could improve the quantitative analysis accuracy of SENLIBS by increasing the sample preparation repeatability.

References