Electroanalysis · 2018 · 24 citations · 26 references
PharmacyBioengineeringAnalytical ChemistryBiosensorsElectroanalytical MethodsBioanalytical ChemistryBioanalytical Method ValidationCancer BiomarkersElectrochemical BiosensorsElectrochemical DetectionAmperometric BiosensorsAmperometric DetectionAmperometric SensingCarbon ElectrodesCarbon-based ElectrodesConcentrationElectrochemical AssaysElectrochemical Sensing
Abstract Flow injection analysis with amperometric detection (FIA‐AD) at screen‐printed carbon electrodes (SPCEs) in optimum medium of Britton‐Robinson buffer (0.04 mol ⋅ L −1 , pH 2.0) was used for the determination of three tumor biomarkers (homovanillic acid (HVA), vanillylmandelic acid (VMA), and 5‐hydroxyindole‐3‐acetic acid (5‐HIAA)). Dependences of the peak current on the concentration of biomarkers were linear in the whole tested concentration range from 0.05 to 100 μmol ⋅ L −1 , with limits of detection ( LOD s) of 0.065 μmol ⋅ L −1 for HVA, 0.053 μmol ⋅ L −1 for VMA, and 0.033 μmol ⋅ L −1 for 5‐HIAA (calculated from peak heights), and 0.024 μmol ⋅ L −1 for HVA, 0.020 μmol ⋅ L −1 for VMA, and 0.012 μmol ⋅ L −1 for 5‐HIAA (calculated from peak areas), respectively.
26
Günther Sperk · Journal of Neurochemistry · 1982 · 139 citations
Charles Saller, Andre I. Salama · Journal of Chromatography B Biomedical Sciences and Applications · 1984 · 88 citations
Luana Lionetto, Alfonso M. Lostia, Antonio Stigliano et al. · Clinica Chimica Acta · 2008 · 79 citations
Medicine, Endocrinology, Oncology +11
Michael H. Joseph, Bharat V. Kadam, D Risby · Journal of Chromatography B Biomedical Sciences and Applications · 1981 · 68 citations
Wenguang He, Yu Yan, Rong Cai et al. · Oncotarget · 2017 · 68 citations · Full text