Effects of Collection and Processing Procedures on Plasma Circulating Cell-Free DNA from Cancer Patients

Bente Risberg, Dana W.Y. Tsui, Heather Biggs, Andrea Ruiz-Valdepenas Martin de Almagro, Sarah‐Jane Dawson, Charlotte Hodgkin, Linda Jones, Christine Parkinson, Anna Piskorz, Francesco Marass,

Journal of Molecular Diagnostics · 2018 · 113 citations · 23 references

DOIFull text

Open access

Concepts

Abstract

Circulating tumor DNA (ctDNA) offers new opportunities for noninvasive cancer management. Detecting ctDNA in plasma is challenging because it constitutes only a minor fraction of the total cell-free DNA (cfDNA). Pre-analytical factors affect cfDNA levels contributed from leukocyte lysis, hence the ability to detect low-frequency mutant alleles. This study investigates the effects of the delay in processing, storage temperatures, different blood collection tubes, centrifugation protocols, and sample shipment on cfDNA levels. Peripheral blood (n = 231) from cancer patients (n = 62) were collected into K3EDTA or Cell-free DNA BCT tubes and analyzed by digital PCR, targeted amplicon, or shallow whole-genome sequencing. To assess pre-analytic effects, plasma was processed under different conditions after 0, 6, 24, 48, 96 hours, and 1 week at room temperature or 4°C, or using different centrifugation protocols. Digital PCR showed that cfDNA levels increased gradually with time in K3EDTA tubes, but were stable in BCT tubes. K3EDTA samples stored at 4°C showed less variation than room temperature storage, but levels were elevated compared with BCT. A second centrifugation at 3000 × g gave similar cfDNA yields compared with higher-speed centrifugation. Next-generation sequencing showed negligible differences in background error or copy number changes between K3EDTA and BCT, or following shipment in BCT. This study provides insights into the effects of sample processing on ctDNA analysis. Circulating tumor DNA (ctDNA) offers new opportunities for noninvasive cancer management. Detecting ctDNA in plasma is challenging because it constitutes only a minor fraction of the total cell-free DNA (cfDNA). Pre-analytical factors affect cfDNA levels contributed from leukocyte lysis, hence the ability to detect low-frequency mutant alleles. This study investigates the effects of the delay in processing, storage temperatures, different blood collection tubes, centrifugation protocols, and sample shipment on cfDNA levels. Peripheral blood (n = 231) from cancer patients (n = 62) were collected into K3EDTA or Cell-free DNA BCT tubes and analyzed by digital PCR, targeted amplicon, or shallow whole-genome sequencing. To assess pre-analytic effects, plasma was processed under different conditions after 0, 6, 24, 48, 96 hours, and 1 week at room temperature or 4°C, or using different centrifugation protocols. Digital PCR showed that cfDNA levels increased gradually with time in K3EDTA tubes, but were stable in BCT tubes. K3EDTA samples stored at 4°C showed less variation than room temperature storage, but levels were elevated compared with BCT. A second centrifugation at 3000 × g gave similar cfDNA yields compared with higher-speed centrifugation. Next-generation sequencing showed negligible differences in background error or copy number changes between K3EDTA and BCT, or following shipment in BCT. This study provides insights into the effects of sample processing on ctDNA analysis. Circulating tumor DNA (ctDNA) in plasma offers new opportunities for noninvasive cancer management. Recent studies have demonstrated its potential for molecular stratification, monitoring tumor response, identifying resistance mutations, and patients at risk of relapse.1Wan J.C.M. Massie C. Garcia-Corbacho J. Mouliere F. Brenton J.D. Caldas C. Pacey S. Baird R. Rosenfeld N. Liquid biopsies come of age: towards implementation of circulating tumour DNA.Nat Rev Cancer. 2017; 17: 223-238Crossref PubMed Scopus (1336) Google Scholar, 2Siravegna G. Marsoni S. Siena S. Bardelli A. Integrating liquid biopsies into the management of cancer.Nat Rev Clin Oncol. 2017; 14: 531-548Crossref PubMed Scopus (1114) Google Scholar Detecting ctDNA in plasma is challenging because it constitutes only a minor fraction of the total cell-free DNA (cfDNA), particularly in early-stage cancers and in the minimal residual disease setting.3Garcia-Murillas I. Schiavon G. Weigelt B. Ng C. Hrebien S. Cutts R.J. Cheang M. Osin P. Nerukar A. Kozarewa I. Garrido J.A. Dowsett M. Reis-Filho J.S. Smith I.E. Turner N.C. Mutation tracking in circulating tumor DNA predicts relapse in early breast cancer.Sci Transl Med. 2015; 7: 302ra133Crossref PubMed Scopus (755) Google Scholar, 4Bettegowda C. Sausen M. Leary R.J. Kinde I. Wang Y. Agrawal N. et al.Detection of circulating tumor DNA in early- and late-stage human malignancies.Sci Transl Med. 2014; 6: 224ra24Crossref PubMed Scopus (2991) Google Scholar A proportion of background wild-type DNA is believed to originate from lysis of white blood cells.5Lui Y.Y. Chik K.W. Chiu R.W. Ho C.Y. Lam C.W. Lo Y.M. Predominant hematopoietic origin of cell-free DNA in plasma and serum after sex-mismatched bone marrow transplantation.Clin Chem. 2002; 48: 421-427Crossref PubMed Scopus (393) Google Scholar Previous studies have highlighted the pre-analytic effects of different processing and collection protocols on plasma ctDNA levels from cancer patients and pregnant women.6El Messaoudi S. Rolet F. Mouliere F. Thierry A.R. Circulating cell free DNA: preanalytical considerations.Clin Chim Acta. 2013; 424: 222-230Crossref PubMed Scopus (402) Google Scholar, 7Chiu R.W.K. Poon L.L.M. Lau T.K. Leung T.N. Wong E.M. Lo Y.M.D. Effects of blood-processing protocols on fetal and total DNA quantification in maternal plasma.Clin Chem. 2001; 47: 1607-1613Crossref PubMed Scopus (303) Google Scholar, 8Barrett A.N. Zimmermann B.G. Wang D. Holloway A. Chitty L.S. Implementing prenatal diagnosis based on cell-free fetal DNA: accurate identification of factors affecting fetal DNA yield.PLoS One. 2011; 6: e25202Crossref PubMed Scopus (93) Google Scholar, 9Rothwell D.G. Smith N. Morris D. Leong H.S. Li Y. Hollebecque A. Ayub M. Carter L. Antonello J. Franklin L. Miller C. Blackhall F. Dive C. Brady G. Genetic profiling of tumours using both circulating free DNA and circulating tumour cells isolated from the same preserved whole blood sample.Mol Oncol. 2016; 10: 566-574Crossref PubMed Scopus (56) Google Scholar On the basis of these results, it is recommended to process whole-blood samples for retrieval of plasma as soon as possible after collection, before in vitro cell lysis. At the same time, a double-centrifugation protocol has been recommended to obtain cell-free plasma, using an initial slow centrifugation speed to separate plasma, then fast centrifugation to clear cellular material.7Chiu R.W.K. Poon L.L.M. Lau T.K. Leung T.N. Wong E.M. Lo Y.M.D. Effects of blood-processing protocols on fetal and total DNA quantification in maternal plasma.Clin Chem. 2001; 47: 1607-1613Crossref PubMed Scopus (303) Google Scholar However, some of these procedures may be difficult to perform in a clinical setting due to lack of appropriate personnel or equipment. To circumvent this, cell-stabilizing blood collection tubes have become available to stabilize cfDNA, enabling a delay in processing, which may be done under more controlled conditions and within centralized laboratories. This study performed a systematic comparison of the effects of different processing protocols and collection tubes on the levels of cfDNA and ctDNA from cancer patients using digital PCR (dPCR). With the growing use of next-generation sequencing (NGS) for the analysis of ctDNA, the effect of different protocols and collection tubes on the performance of targeted amplicon and shallow whole-genome sequencing (sWGS) for quantification of plasma DNA was also investigated. The study was designed to include five different modules: Module 1 investigated the effects of delayed processing on the levels of circulating DNA (cfDNA and ctDNA) in plasma collected in K3EDTA tubes (9 mL S-Monovette; Sarstedt, Nümbrecht, Germany). The separation of plasma was delayed for different durations: 0, 6, 24, 48, and 96 hours, and 1 week at room temperature (19°C to 25°C). Module 2 investigated the effects of storage temperature on the levels of circulating DNA in plasma collected in K3EDTA tubes. Samples were stored at room temperature or at 4°C before processing at the following hours post-collection: 0, 24, 48, and 96 hours. Module 3 investigated the effects of collection devices on the levels of circulating DNA. Blood samples from each patient were collected at the same time point into K3EDTA tubes and cell-stabilization blood collection tubes (10 mL Cell-Free DNA BCT; Streck, La Vista, NE), respectively. BCTs contain a proprietary formaldehyde-free preservative that stabilizes nucleated blood cells preventing the release of genomic DNA.10Fernando M.R. Chen K. Norton S. Krzyzanowski G. Bourne D. Hunsley B. Ryan W.L. Bassett C. A new methodology to preserve the original proportion and integrity of cell-free fetal DNA in maternal plasma during sample processing and storage.Prenat Diagn. 2010; 30: 418-424Crossref PubMed Scopus (74) Google Scholar, 11Norton S.E. Lechner J.M. Williams T. Fernando M.R. A stabilizing reagent prevents cell-free DNA contamination by cellular DNA in plasma during blood sample storage and shipping as determined by digital PCR.Clin Biochem. 2013; 46: 1561-1565Crossref PubMed Scopus (136) Google Scholar The samples were processed at the following times post-collection: 0, 96 hours, and 1 week. Module 4 investigated the effects of different centrifugation protocols on the levels of circulating DNA. Module 5 investigated the effects of shipment on samples collected in BCT tubes at ambient temperature. For modules 1, 2, 3, and 5, plasma was separated from blood using a double-centrifugation protocol (protocol A): a first centrifugation at 820 × g for 10 minutes in a mega-centrifuge (Thermo Sorvall Legend RT; Thermo Fisher Scientific, Waltham, MA), then subjected to a second centrifugation step of the plasma supernatant at 14,000 × g for 10 minutes in a benchtop micro-centrifuge (Heraeus Fresco 21; Thermo Fisher Scientific). For module 4, blood aliquots from the same patients were processed with three different protocols: protocol A as above, protocol B with the first centrifugation performed at 1600 × g and the second centrifugation at 14,000 × g for 10 minutes in a bench top micro-centrifuge, and protocol C with both first and second centrifugations performed in the same mega-centrifuge, initially at 1600 × g for 10 minutes, then at 3000 × g for 10 minutes. Peripheral whole blood was collected from 62 patients in total: 47 patients with high-grade serous ovarian cancer and 15 patients with metastatic breast cancer. Informed consent was from each patient with protocols by an to mL blood from each patient was processed to each analysis DNA from module 5, was from an mL to plasma using the Circulating to the that of was DNA was the to A PCR was to each sample as an to assess M. D. T. S. S. J. D. M. D. Caldas C. Rosenfeld N. of circulating tumor DNA to metastatic breast Med. 2013; PubMed Scopus Google Scholar module 5, DNA was from plasma on a using a DNA was stored at analysis. A total of blood samples aliquots were analyzed in 1 the number of plasma samples collected for each that the collection was designed in a that each sample from processing collection delayed processing a sample that was collected in K3EDTA and processed using centrifugation protocol and was as the sample for each The levels of circulating DNA cfDNA or were as a of the with the sample collected under the same processing conditions be to the effect of the processing were collected from different A more of the of samples in each module is in of for before sample The effects of delayed The effects of storage temperature. Samples were stored both in room temperature and at that tubes were stored at room temperature and at 4°C, and The effects of collection devices The effects of different centrifugation The effects of shipment in in a new Module The effects of delayed Module The effects of storage temperature. Samples were stored both in room temperature and at that tubes were stored at room temperature and at 4°C, and Module The effects of collection devices Module The effects of different centrifugation Module The effects of shipment in BCT. samples from ovarian and breast cancer patients were first by the as T. M. C. D. F. M. D. J. Caldas C. Brenton J.D. Rosenfeld N. identification and monitoring of cancer by targeted sequencing of plasma Transl Med. PubMed Scopus Google Scholar using an that a amplicon in a in the to cfDNA M. D. T. S. S. J. D. M. D. Caldas C. Rosenfeld N. of circulating tumor DNA to metastatic breast Med. 2013; PubMed Scopus Google Scholar, J. R. A. A. in the cancer with digital PCR Chem. 2010; PubMed Scopus Google Scholar ctDNA levels were then determined by using designed to mutant and wild-type in or or in or A of the samples analyzed is in and of and amplicon and conditions in in The levels of cfDNA and ctDNA were from the number of a and were to the number of to of DNA into the The total number of of DNA mL of plasma were into the fraction of the DNA and the proportion of sample during the process the The levels of ctDNA were as mutant fraction the fraction of mutant DNA by the total cfDNA as a or as mutant For the of different protocols in the the at each processing as a from the sample that was collected in K3EDTA and processed to protocol To the effects of different collection devices and processing protocols on the performance of plasma samples from modules were analyzed by sequencing as T. M. C. D. F. M. D. J. Caldas C. Brenton J.D. Rosenfeld N. identification and monitoring of cancer by targeted sequencing of plasma Transl Med. PubMed Scopus Google Scholar is a targeted amplicon sequencing that identification and quantification of low-frequency mutant in plasma genomic was performed using an to an of than sequencing were and as T. M. C. D. F. M. D. J. Caldas C. Brenton J.D. Rosenfeld N. identification and monitoring of cancer by targeted sequencing of plasma Transl Med. PubMed Scopus Google Scholar To assess the effect of collection and processing procedures on the background error during the was at each within A and for for Scholar from the using the and R. S. M. S. L. T. R. F. M. J. A. P. D. L. M. genomic analysis with 2015; PubMed Scopus Google Scholar as by the or the were To the effects of shipping on copy number samples in module 5 were also subjected to P. J. S. F. K. T. M. C. S. M. F. M. S. K. G. M.R. copy number changes in the of patients with cancer whole-genome Med. 2013; PubMed Scopus Google Scholar a DNA was from 2 to 10 of cfDNA from each sample using the and on an to using sequencing. were analyzed using a that the were to the human using R. and accurate with PubMed Scopus Google Scholar after B. A. T. J. N. G. G. R. The and PubMed Scopus Google Scholar, A for and from sequencing 2011; PubMed Scopus Google Scholar was to to PCR and were using and these were from analysis with of and in each sample were to 3 to have similar between patients and copy number analysis was performed in A and for for Scholar using the a that for copy number analysis of were into the of the in each were to for and and to and changes were from analysis a of were using both the and and an was The in circulating DNA levels between different in each module was analyzed using and was To assess the of to the of the of differences were between copy This provides a of the of the sample that is less on copy number variation and more on of the of and on Scholar To the three collection in were between the copy number of the three collection a was on these to the of the copy number between samples (n = were collected in K3EDTA tubes. from each collection was processed The tubes were stored at room temperature and processed at different time 6, 24, 48, 96 hours, and 1 week. by showed that the levels of cfDNA in the plasma samples increased gradually with delay in the processing the fraction of ctDNA the levels of cfDNA increased after 48, 96 hours, and 1 week of the mutant fraction of ctDNA after 96 hours and 1 week of delay Previous have that in analysis of circulating cell-free DNA from maternal plasma, changes in total cfDNA, the levels of fetal DNA stable in different storage and processing A.N. Zimmermann B.G. Wang D. Holloway A. Chitty L.S. Implementing prenatal diagnosis based on cell-free fetal DNA: accurate identification of factors affecting fetal DNA yield.PLoS One. 2011; 6: e25202Crossref PubMed Scopus (93) Google Scholar, D. S. R. G. D. M. blood collection, and storage conditions for cell free DNA to prenatal Biochem. 2013; 46: PubMed Scopus Google Scholar that the of mutant as of plasma, were stable the different processing time with compared to samples that were processed and samples (n = were collected in K3EDTA tubes and processed to plasma or after 24, 48, and 96 hours. The tubes were stored in at room temperature (19°C to or at at room showed that the levels of cfDNA increased after hours. at 4°C, the levels increased after hours but were than at room temperature delayed for 96 hours, samples at room temperature and 4°C increased The changes in mutant fraction showed an similar the of available were for analysis K3EDTA for each collection was processed and as a sample (n = The K3EDTA tubes were stored for 96 hours (n = and 1 week (n = at room temperature. Cell-free DNA were stored at room temperature and processed (n = or delayed for 96 hours (n = and 1 week (n = The cfDNA levels increased after 1 week in K3EDTA tubes, but at similar levels in BCT The changes in the mutant fraction showed an similar but the of available were for analysis The mutant fraction from patients that were collected in K3EDTA and processed the samples that were collected in BCT was compared and processed after 1 The levels of ctDNA were similar for patients but for patient samples was in the of mutant plasma between storage in the The effects of collection and processing procedures on the background error during analysis were using targeted amplicon sequencing. different with different error T. M. C. D. F. M. D. J. Caldas C. Brenton J.D. Rosenfeld N. identification and monitoring of cancer by targeted sequencing of plasma Transl Med. PubMed Scopus Google Scholar The of the of was as to was using different collection devices and processing conditions samples (n = were collected in K3EDTA tubes and processed from the same patients were processed using three different centrifugation protocols to as in and were differences the three protocols on the total circulating DNA levels as by 4, A and or in mutant fraction as by targeted amplicon sequencing 4, C and three tubes of blood were from each patient (n = K3EDTA tubes were processed cell-free DNA BCT was collected and stored at room temperature within the same centralized processing the BCT was and to the same from were and processed within hours from the time of BCTs were processed after 96 hours and was processed after 5 The stored BCTs were processed at the same time as the was in cfDNA levels between the three collection 5, A and were by in and were differences in mutant fraction using the different collection 5, C and To the effects of collection on copy number analysis was performed on patients with and from patient were from analysis because the total for of the collection was 1 This is the recommended for shallow A. Y. P. S. for cancer and tumour cell of copy number from next-generation PubMed Scopus Google Scholar The copy number the three collection were an of to and The differences in copy number an of the copy number in plasma samples processed with and The same and in were in three protocols. the of in the using that were the of the of patients showed similar levels between the different tubes and protocols. studies have demonstrated the potential of using plasma as a for noninvasive cancer management. is in ctDNA as a liquid in both clinical and the of mutant in plasma may be particularly in early-stage it is to and pre-analytic sample processing procedures to the of samples for accurate quantification of mutant the pre-analytic effects of blood sample processing the use of different blood collection tubes, storage and centrifugation were on analysis of cfDNA using different molecular targeted amplicon, and sequencing. that levels of cfDNA stable in K3EDTA tubes at room temperature for to hours. delayed hours, storage of K3EDTA blood at 4°C to delay the in background is that a study demonstrated that the samples in tubes at 4°C the cfDNA levels stable for a of 3 S. B. M. L. R. A. J. J. N. I. D. Sausen M. The effect of preservative and temperature on the analysis of circulating tumor 2017; PubMed Scopus Google Scholar This with the that K3EDTA tubes at 4°C the of cfDNA compared with room temperature collection into cell-free DNA BCT tubes at room temperature stable cfDNA levels for at a week. tubes delayed and centralized blood processing, with delayed plasma have cell-stabilization tubes as and Blood tubes and demonstrated similar sample processing was D.G. Smith N. Morris D. Leong H.S. Li Y. Hollebecque A. Ayub M. Carter L. Antonello J. Franklin L. Miller C. Blackhall F. Dive C. Brady G. Genetic profiling of tumours using both circulating free DNA and circulating tumour cells isolated from the same preserved whole blood sample.Mol Oncol. 2016; 10: 566-574Crossref PubMed Scopus (56) Google Scholar, K. A. of BCT and blood collection tubes for cell-free circulating DNA studies in 2017; PubMed Scopus Google Scholar cell-free tubes have become available Cell-free DNA Blood and Blood tubes, and it be to these to assess performance for sample processing procedures before next-generation sequencing and analysis of have a of the in the sample processing in a For in the processing may be delayed due to of to processing, or collection hours. some clinical have to the of with the second centrifugation to perform the recommended double-centrifugation The ability to delay processing by into cell-stabilization tubes, or the to perform the centrifugation in a of different of or at 4°C after collection for a the of For samples collected a shipment may be before processing to pre-analytic factors and This study showed in background with or However, studies have that the shipping temperature of cell-free DNA BCT was to be a to of for ctDNA I. A. J. F. F. of cfDNA blood collection tubes for liquid One. 2016; PubMed Scopus Google Scholar these were at temperatures, at and which the cellular in an elevated of genomic DNA and a in plasma studies that shipment temperature be controlled by the use of or temperature devices to Previous studies have on analysis using PCR or that at a With an number of molecular profiling have been using N. and the of clinical Rev 2013; 14: PubMed Scopus Google Scholar which provides a and genomic than a is to also the effects of cfDNA sample processing on the performance of analysis. is particularly to using a collection a preservative has the potential to DNA which may be as genomic A study the of sample collection in tubes on the analysis of copy number using and differences in compared with D.G. Smith N. Morris D. Leong H.S. Li Y. Hollebecque A. Ayub M. Carter L. Antonello J. Franklin L. Miller C. Blackhall F. Dive C. Brady G. Genetic profiling of tumours using both circulating free DNA and circulating tumour cells isolated from the same preserved whole blood sample.Mol Oncol. 2016; 10: 566-574Crossref PubMed Scopus (56) Google Scholar study with BCT and K3EDTA tubes, the effects of processing on the background error during targeted amplicon sequencing and were different error were in different but was in background error of the of collection and sample processing The analysis with in that copy number were J. J. The in vitro of circulating tumour 2016; PubMed Scopus Google Scholar of these insights for the potential of in molecular the analysis of ctDNA, it is to also the of factors on or genomic as or DNA some of which have been K. N. L. K. A. L. J. J.A. of plasma processing on and analysis of circulating One. 2013; PubMed Scopus Google Scholar but more is quantification be by the levels of total or DNA that is from the blood is to the effects of sample processing procedures be in a similar to the effects on circulating DNA. With the of genomic and targeted the for a molecular profiling is cell-free a for during of cancer have to the of for cell-free DNA analysis. the of different pre-analytic factors the process and studies to for clinical of circulating tumor DNA and its into clinical the patients for to the at which is by the for for and the of the for sequencing and D.G. and designed the and processed collected clinical and and D.G. performed and analyzed and D.G. the the with of circulating tumor DNA (ctDNA) as plasma, in different storage and processing conditions and the from the sample collected in K3EDTA tubes and processed with of samples from collected in K3EDTA tubes and processed collected in BCT and processed after 1 week of with comparison between copy number of plasma samples processed with different protocols. the sample was collected in K3EDTA and processed BCT the sample was collected in BCT and in the and processed BCT the sample was collected in BCT and stored at room temperature processing with the The copy number were as the of the The were by and the on top of each with effects of shipping using cell-free DNA BCT on copy number from whole-genome sequencing of an ovarian cancer plasma sample processed with three different protocols: collection in K3EDTA and processed collection in BCT, stored at room processed the sample collection in BCT, in the and processed The were as the of the The genomic of similar copy which the of the and a a with from the copy number by of differences between copy number for the different protocols. The to patient with with with with

References

23