Journal of neurosurgery · 2018 · 15 citations · 12 references
MedicineNeurologyNeuroscienceBioinstrumentationClinical NeuroscienceSensorsIntracranial Pressure MonitoringPostoperative MonitoringContinuous Intracranial Pressure MonitoringDevice DriftIntracranial Pressure Measurement AccuracyIntracranial Pressure Monitoring DevicesLong-term Device ReliabilityLong-term Implant ReliabilityNoninvasive Intracranial Pressure EstimationNoninvasive Intracranial Pressure Monitoring
Intracranial pressure (ICP) measurements are imperative for the proper diagnosis and treatment of several neurological disorders. Telemetric sensors have shown their utility for ICP estimation in short-term monitoring in humans. However, their long-term reliability is uncertain. The authors present the case of a 37-year-old woman diagnosed with benign intracranial hypertension and obesity. The patient underwent gastric bypass surgery for ICP control. In order to monitor ICP before and after bariatric surgery, a Neurovent-P-tel sensor was implanted in the left frontal lobe. After gastric bypass, normal ICP values were recorded, and the patient's visual fields improved. However, the patient experienced incapacitating daily headaches. The authors decided to implant a Codman Microsensor ICP transducer in the right frontal lobe to assess the long-term reliability of the Neurovent-P-tel measurements. A comparison of the recordings at 24 and 48 hours showed good correlation and reliability during long-term monitoring with the Neurovent-P-tel, with minimal zero drift after 11 months of implantation.
12
Gastric Surgery for Pseudotumor Cerebri Associated With Severe Obesity
Harvey J. Sugerman, Warren L. Felton, Aristides Sismanis et al. · Annals of Surgery · 1999 · 231 citations · Full text
Medicine, Surgery, Anesthesia +8
Telemetric Intracranial Pressure Monitoring with the Raumedic Neurovent P-tel
Sebastian Antes, Christoph A. Tschan, Michael Heckelmann et al. · World Neurosurgery · 2016 · 71 citations
Medicine, Neurology, Neuroscience +12
ASSESSMENT OF ZERO DRIFT IN THE CODMAN INTRACRANIAL PRESSURE MONITOR
Yahia Al-Tamimi, Adel Helmy, Seb Bavetta et al. · Neurosurgery · 2009 · 43 citations