Natural SciencesBiomedical EngineeringBiophysicsBiophotonicsPhotonicsSpectroscopyDetector TechnologyOptical DetectorsOptical SpectroscopyX-ray DetectionX-ray DetectorsX-ray SpectroscopyBrownian DynamicsContinuous Mode OperationDynamic RangeFrame AcquisitionPhoton Counting DetectionPhoton-counting DetectorsReadoutTime Scales
Small-angle X-ray photon correlation spectroscopy (XPCS) measurements spanning delay times from 826 ns to 52.8 s were performed using a photon-counting pixel array detector with a dynamic range of 0–3 (2 bits). Fine resolution and a wide dynamic range of time scales was achieved by combining two modes of operation of the detector: (i) continuous mode, where data acquisition and data readout are performed in parallel with a frame acquisition time of 19.36 µs, and (ii) burst mode, where 12 frames are acquired with frame integration times of either 2.56 µs frame −1 or 826 ns frame −1 followed by 3.49 ms or 1.16 ms, respectively, for readout. The applicability of the detector for performing multi-speckle XPCS was demonstrated by measuring the Brownian dynamics of 10 nm-radius gold and 57 nm-radius silica colloids in water at room temperature. In addition, the capability of the detector to faithfully record one- and two-photon counts was examined by comparing the statistical distribution of photon counts with expected probabilities from the negative binomial distribution. It was found that in burst mode the ratio of 2 s to 1 s is markedly smaller than predicted and that this is attributable to pixel-response dead-time.
47
Ch. Broennimann, Eric F. Eikenberry, B. Henrich et al. · Journal of Synchrotron Radiation · 2006 · 482 citations · Full text
Performance of single-photon-counting PILATUS detector modules
Philipp Kraft, A. Bergamaschi, Ch. Broennimann et al. · Journal of Synchrotron Radiation · 2009 · 404 citations · Full text