Analytical Chemistry · 2018 · 80 citations · 62 references
Natural SciencesBiochemistryChemistryAnalytical ChemistryMass Spectrometric AnalysisProteomicsMass AnalyzersMass SpectrometryPeptide AnalysisProtein Complex AnalysisCollision Cross SectionsFourier Transform Mass SpectrometryInstrument Set ConfigurationIon Mobility SpectrometryNative Mass SpectrometryNew Instrument DesignProtein ComplexesProtein Mass Spectrometry
A new instrument configuration for native ion mobility-mass spectrometry (IM-MS) is described. Macromolecule ions are generated by using a static ESI source coupled to an RF ion funnel, and these ions are then mobility and mass analyzed using a periodic focusing drift tube IM analyzer and an Orbitrap mass spectrometer. The instrument design retains the capabilities for first-principles determination of rotationally averaged ion-neutral collision cross sections and high-resolution measurements in both mobility and mass analysis modes for intact protein complexes. Operation in the IM mode utilizes FT-IMS modes (originally described by Knorr ( Knorr , F. J. Anal. Chem . 1985 , 57 ( 2 ), 402 - 406 )), which provides a means to overcome the inherent duty cycle mismatch for drift tube (DT)-IM and Orbitrap mass analysis. The performance of the native ESI-FT-DT-IM-Orbitrap MS instrument was evaluated using the protein complexes Gln K (MW 44 kDa) and streptavidin (MW 53 kDa) bound to small molecules (ADP and biotin, respectively) and transthyretin (MW 56 kDa) bound to thyroxine and zinc.
62
The Orbitrap: a new mass spectrometer
Qizhi Hu, Robert J. Noll, Hongyan Li et al. · Journal of Mass Spectrometry · 2005 · 1.3K citations
How many human proteoforms are there?
Ruedi Aebersold, Jeffrey N. Agar, I. Jonathan Amster et al. · Nature Chemical Biology · 2018 · 944 citations · Full text
Membrane proteins bind lipids selectively to modulate their structure and function
Arthur Laganowsky, Eamonn Reading, Timothy M. Allison et al. · Nature · 2014 · 810 citations · Full text
Steven Pringle, Kevin Giles, Jason Wildgoose et al. · International Journal of Mass Spectrometry · 2006 · 808 citations