Applied Optics · 2018 · 56 citations · 37 references
Natural SciencesPhysicsSystems EngineeringInstrumentationOpticsPhotonicsCold Atom PhysicsInterferometryLaser PhysicsLaser TrappingOptical InstrumentationAtom Interferometry GravimetryLaser Interferometric GravimetryMagneto-optical EffectsMagneto-optical TrappingMagneto-optical Traps
We have designed and realized a compact portable laser system for high-sensitivity mobile cold atom interferometers. The laser system is mounted on a single module with dimensions of 45  cm×45  cm×16  cm and emits lights directly on up to 13 fiber ports for a two-dimensional magneto-optical trap, atom fountain, Raman transition, and normalized detection. A double-sided optical structure and mounts without kinematic adjustment are designed to achieve high-level integration and stability. The laser system is applied to a mobile atom gravimeter and achieves a sensitivity of 28  μGal/Hz1/2. Experiencing 1200-km-long truck transportation, the laser system could be restored to good operating status without internal realignment.
37
Trapping of Neutral Sodium Atoms with Radiation Pressure
E. L. Raab, Mara Prentiss, Alex Cable et al. · Physical Review Letters · 1987 · 1.5K citations
Atomic interferometry using stimulated Raman transitions
Mark A. Kasevich, Steven Chu · Physical Review Letters · 1991 · 1.3K citations
High-precision gravity measurements using atom interferometry
Achim Peters, Keng Yeow Chung, Steven Chu · Metrologia · 2001 · 939 citations
Measurement of the fine-structure constant as a test of the Standard Model