Platelet Dysfunction in Cardiopulmonary Bypass: An Experimental Comparative Study between a Centrifugal and a New Pulsatile Pump

Jesús Herreros, Enrique Berjano, Pedro Más, Clemente Padrós, Laura Sales‐Nebot, Wouter Vlaanderen, Pedro Bedate Díaz, José A. Páramo, Gregorio Rábago, Salvador Mercé

The International Journal of Artificial Organs · 2003 · 14 citations · 32 references

Concepts

Abstract

The aim of this investigation was to study the effect of a new pulsatile pump for extracorporeal circulation (ECC) on platelet count and platelet function with respect to a Biomedicus centrifugal pump. Thirteen pigs, 8 in the pulsatile group (PG) and 5 in the centrifugal group (CG), underwent a partial extracorporeal circulation lasting 3 h. The animals were sacrificed 3 h post-ECC. The platelet study was both quantitative (platelet count) and qualitative (platelet function analysis) by assessing the closure time (CT) with a PFA-100 system. The decrease in platelet number from basal to 3 h post ECC was only significant in CG (p = 0.009). The platelet function was impaired in both groups, but the value of CT with col/ADP increased significantly only in CG (p < 0.001). The increase of CT with col/EPI was greater in CG (p = 0.07) than in PG (p = 0.2). The results indicated that the new pulsatile pump preserves platelets quantitatively and qualitatively well compared to a Biomedicus pump.