Health ProfessionsOtolaryngologySurgeryCochlear ImplantsHearing ScienceRadiographic AnatomyEar AnatomyImplant Position VerificationSurgical ImplantationTemporal Bone AnatomyTemporal Bone SurgeryThree-dimensional ReconstructionBone Conduction Hearing ImplantationBone Conduction Transducer PlacementSurgical Access PlanningSurgical PlanningTemporal Bone Computed TomographyTemporal Bone Pathway AssessmentTemporal Bone SegmentsTransducer Design
SUMMARY: ) (BB), focusing on the identification of the proper location and placement of the transducer. 3D printed (3DP) models of three human cadaveric temporal bones, previously submitted to CT scan, were created with the representation of a topographic bone thickness map and the sinus pathway on the outer surface. The 3DP model was used to detect the most suitable location for the BB. A 3DP transparent mask that faithfully reproduced the surface of both the temporal bone and the 3DP model was also developed to correctly transfer the designated BB area. The accuracy of the procedure was verified by CT scan: a radiological marker was used to evaluate the degree of correspondence of the transducer site between the 3DP model and the human temporal bone. The BB positioning was successfully performed on all human temporal bones, with no difficulties in finding the proper location of the transducer. A mean error of 0.13 mm was found when the transducer site of the 3DP model was compared to that of the human temporal bone. The employment of 3D printing technology in surgical planning of BB positioning showed feasible results. Further studies will be required to evaluate its clinical applicability.