Scientific Reports · 2018 · 20 citations · 54 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsPhotonicsSpectroscopyBiological SensorsBiomedical InstrumentationBiomedical SensorsOptical BiosensingOptical BiosensorsOptical SpectroscopyControlled ExperimentsCoplanar Waveguide MatchingCoplanar WaveguidesExperimental ValidationNumerical ModelingOptical DetectionSpatial InhomogeneitySpectroscopic Imaging
Abstract The analytical and numerical design, implementation, and experimental validation of a new grounded closed coplanar waveguide for wide-band electromagnetic exposures of cells and their optical detection in real-time is reported. The realized device fulfills high-quality requirements for novel bioelectromagnetic experiments, involving elevated temporal and spatial resolutions. Excellent performances in terms of matching bandwidth (less than −10 dB up to at least 3 GHz), emission (below 1 × 10 −6 W/m 2 ) and efficiency (around 1) have been obtained as revealed by both numerical simulations and experimental measurements. A low spatial electric field inhomogeneity (coefficient of variation of around 10 %) has been achieved within the cell solutions filling the polydimethylsiloxane reservoir of the conceived device. This original bio-chip based on the grounded closed coplanar waveguide concept opens new possibilities for the development of controlled experiments combining electromagnetic exposures and sophisticated imaging using optical spectroscopic techniques.
54