Journal of Clinical Microbiology · 2018 · 62 citations · 7 references
Life SciencesInfectious DiseasesMicrobiologyBacteriologyDiagnostic MicrobiologyMedical MicrobiologyCulture TechniquesUrinary Tract Infection DiagnosticsUrinary Tract InfectionsBacteriuria DetectionGram-negative Bacterial InfectionsRapid Screening TestsRapid Urine DiagnosticsUrinary Tract Infection Antibiotic TherapyUrine Culture Interpretation
CFU/ml monobacterial samples, the sensitivity and specificity for reporting GN bacteria were 91.7 and 90.0%, respectively. In summary, UF-5000 demonstrated potential utility for the rapid screening of negative bacterial cultures. However, this utility is dependent on the patient population; cutoff optimizations must be performed for specific populations. In addition, UF-5000 presented improved performance in characterizing GP and GN bacteria, although the concurrence rates were not high enough to replace routine cultures.
7