Rapid Screening of Urinary Tract Infection and Discrimination of Gram-Positive and Gram-Negative Bacteria by Automated Flow Cytometric Analysis Using Sysmex UF-5000

Seon Young Kim, Yumi Park, Hyun‐Jin Kim, Jimyung Kim, Sun Hoe Koo, Gye Cheol Kwon

Journal of Clinical Microbiology · 2018 · 62 citations · 7 references

DOIFull text

Open access

Concepts

Abstract

CFU/ml monobacterial samples, the sensitivity and specificity for reporting GN bacteria were 91.7 and 90.0%, respectively. In summary, UF-5000 demonstrated potential utility for the rapid screening of negative bacterial cultures. However, this utility is dependent on the patient population; cutoff optimizations must be performed for specific populations. In addition, UF-5000 presented improved performance in characterizing GP and GN bacteria, although the concurrence rates were not high enough to replace routine cultures.

References