Diagnosis of Bacteriuria and Leukocyturia by Automated Flow Cytometry Compared with Urine Culture

Barbara Pieretti, Pietro Brunati, Beatrice Pini, Carlo Colzani, Pierluigi Congedo, Marco Rocchi, Riccardo Terramocci

Journal of Clinical Microbiology · 2010 · 96 citations · 10 references

DOIFull text

Open access

Concepts

Abstract

Urinary tract infection (UTI) is a widespread disease, and thus, the most common samples tested in diagnostic microbiology laboratories are urine samples. The "gold standard" for diagnosis is still bacterial culture, but a large proportion of samples are negative. Unnecessary culture can be reduced by an effective screening test. We evaluated the performance of a new urine cytometer, the Sysmex UF-1000i (Dasit), on 703 urine samples submitted to our laboratory for culture. We compared bacteria and leukocyte (WBC) counts performed with the Sysmex UF-1000i to CFU-per-milliliter quantification on CPS agar to assess the best cutoff values. Different cutoff values of bacteria/ml and WBC/ml were compared to give the best discrimination. On the basis of the results obtained in this study, we suggest that when the Sysmex UF-1000i analyzer is used as a screening test for UTI the cutoff values should be 65 bacteria/ml and 100 WBC/ml. Diagnostic performance in terms of sensitivity (98.2%), specificity (62.1%), negative predictive value (98.7%), positive predictive value (53.7%), and diagnostic accuracy (73.3%) were satisfactory. Screening with the Sysmex UF-1000i is acceptable for routine use. In our laboratory, we have reduced the number of bacterial cultures by 43%, speeded up their reporting, and decreased the inappropriate use of antibiotics.