Publication | Closed Access
Phantom, clinical, and texture indices evaluation and optimization of a penalized‐likelihood image reconstruction method (Q.Clear) on a <scp>BGO PET</scp>/<scp>CT</scp> scanner
48
Citations
23
References
2018
Year
Our phantom and patients demonstrate that improvement of CNR and SNR of Q.Clear algorithm which depends on the studied conditions and the penalization factor. Using the Q.Clear reconstruction algorithm in a BGO scanner, a β value of 350 and 200 appears to be the optimal value for 18F-FDG oncology and brain PET/CT, respectively.
| Year | Citations | |
|---|---|---|
Page 1
Page 1