IEEE Transactions on Medical Imaging · 2018 · 95 citations · 32 references
EngineeringComputer EngineeringNuclear MedicineDiagnostic ImagingField-programmable Gate ArraysReconfigurable ComputingComputed TomographyField-programmable Gate Array ArchitectureFpga ArchitectureImage ReconstructionMedical ImagingPositron Emission TomographyData ProcessingImage Reconstruction HardwarePet Image ReconstructionPet-computed TomographyProgrammable Logic ArrayProgrammable Logic DesignReal-time Data ProcessingTomographic System Design
A novel approach to tomographic data processing has been developed and evaluated using the Jagiellonian positron emission tomography scanner as an example. We propose a system in which there is no need for powerful, local to the scanner processing facility, capable to reconstruct images on the fly. Instead, we introduce a field programmable gate array system-on-chip platform connected directly to data streams coming from the scanner, which can perform event building, filtering, coincidence search, and region-of-response reconstruction by the programmable logic and visualization by the integrated processors. The platform significantly reduces data volume converting raw data to a list-mode representation, while generating visualization on the fly.
32
GATE: a simulation toolkit for PET and SPECT
S Jan, G. Santin, D. Strul et al. · Physics in Medicine and Biology · 2004 · 2.1K citations · Full text