MedicineRadiation OncologyRadiation DosimetryRadiation MeasurementSpectroscopyAbsorbed DoseDose CalibrationExternal Beam DosimetryFilm DosimetryProton DosimetryRadiation Dose MeasurementDose Range CharacterizationDose ResponseFilm Dosimetry CalibrationGafchromic Film DosimetryOptical Density CalibrationOptical Density MeasurementProton Beam CalibrationProton Beam Dosimetry Calibration
Radiochromic film is used extensively in many medical, industrial, and scientific applications. In particular, the film is used in analysis of proton generation and in high intensity laser-plasma experiments where very high dose levels can be obtained. The present study reports calibration of the dose response of Gafchromic EBT3 and HD-V2 radiochromic films up to high exposure densities. A 2D scanning confocal densitometer system is employed to carry out accurate optical density measurements up to optical density 5 on the exposed films at the peak spectral absorption wavelengths. Various wavelengths from 400 to 740 nm are also scanned to extend the practical dose range of such films by measuring the response at wavelengths removed from the peak response wavelengths. Calibration curves for the optical density versus exposure dose are determined and can be used for quantitative evaluation of measured doses based on the measured optical densities. It was found that blue and UV wavelengths allowed the largest dynamic range though at some trade-off with overall accuracy.