Journal of Magnetic Resonance Imaging · 2018 · 28 citations · 38 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDecay CorrectionMagnetic Resonance ImagingMagnetic Resonance Pulse SequencesMedical ImagingRelaxation MechanismsExponential Curve FittingShapiro - Wilk TestSpin EchoSpin Echo ImagingT2 RelaxationTransverse RelaxationTransverse Relaxation TimeWilcoxon Signed-rank Test
Background Exponential fitting of multiecho spin echo sequences with skipped echoes is still commonly used for quantification of transverse relaxation (T 2 ). Purpose To examine the efficacy of skipped echo methods for T 2 quantification against computational modeling of the exact signal decay. Study Type Prospective comparison of methods. Subjects/Phantom Eight volunteers were imaged at 4.7T, six volunteers at 1.5T, and phantoms ([MnCl 2 ] = 68–270 mM). Field Strength/Sequence 1.5T and 4.7T; multiple‐echo spin echo. Assessment Exponential fitting for T 2 using all echoes, skipping the first echo or skipping all odd echoes, compared with Bloch simulations. Resulting T 2 values were examined over a range of T 2 (10–150 msec), refocusing flip angles (90–270°), and echo train lengths (ETL = 6–32). Statistical Tests Shapiro–Wilk tests and Q‐Q plots were used to check for normality of data. Paired sample t ‐tests and Wilcoxon rank tests were used to compare fitting models using α = 0.05. Multiple comparisons were accounted for with Bonferroni correction. Results In examined regions of interest, typical incorrect estimation of T 2 ranged from 23–39% for exponential fitting of all echoes, or 15–32% for skipped echo methods. In vivo, T 2 estimation error was reduced to as little as 10% with skipped echo methods using 180° refocusing and ETL = 8, although error varied due to refocusing angle, T 2 , and ETL. In vivo, skipped echo T 2 values were significantly different than all echo exponential fitting ( P < 0.004), but also were significantly different from reference values ( P < 0.002, except frontal white matter). Simulations showed skipping the first echo was the most effective form of exponential fitting, in particular for T 2 <50 msec and ETL = 8, with potential to reduce T 2 errors to 10%, depending on refocusing angle and T 2 . Data Conclusion Skipping echoes is insufficient for avoiding stimulated echo contamination. Resulting T 2 errors depend on a complicated interplay of T 2 , refocusing angle, and ETL. Modeling of the multiecho sequence is recommended. Level of Evidence : 2 Technical Efficacy : Stage 1 J. Magn. Reson. Imaging 2018;47:1432–1440.
38
Quantitative MR Imaging of Brain Iron: A Postmortem Validation Study
Christian Langkammer, Nikolaus Krebs, Walter Goessler et al. · Radiology · 2010 · 527 citations
Medicine, Neuroscience, Radiology +15