Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)

John M. Pauly, Patrick Le Roux, Dwight G. Nishimura, Albert Macovski

IEEE Transactions on Medical Imaging · 1991 · 790 citations · 14 references

Concepts

TL;DR

Linear‑phase pulses can be refocused with a gradient reversal and used as spin‑echo pulses. An overview of the Shinnar‑Le Roux (SLR) algorithm is presented. The algorithm enables analytic design of pulses with specified slice profiles, allowing trade‑offs among performance parameters, illustrated by examples of linear‑phase, minimum‑phase, and maximum‑phase pulses. The study demonstrates that SLR pulse performance can be accurately specified analytically, and that while minimum‑ and maximum‑phase pulses achieve superior slice profiles, they cannot be fully refocused.

Abstract

An overview of the Shinnar-Le Roux (SLR) algorithm is presented. It is shown how the performance of SLR pulses can be very accurately specified analytically. This reveals how to design a pulse that produces a specified slice profile and allows the pulse designer to trade off analytically the parameters describing the pulse performance. Several examples are presented to illustrate the more important tradeoffs. These include linear-phase and minimum- and maximum-phase pulses. Linear-phase pulses can be refocused with a gradient reversal and can be used as spin-echo pulses. Minimum- and maximum-phase pulses have better slice profiles, but cannot be completely refocused.