PLoS ONE · 2018 · 58 citations · 32 references
Health ProfessionsOphthalmologyOptometryClinical OptometryOcular ImagingOphthalmic SurgeryCorneal ImagingCorneal TopographyEye ExaminationOcular BiometryCentral Corneal Thickness MeasurementCorneal DiameterIntraclass CorrelationOcular BiometricsPaired T-testPupil DiameterReproducibility AnalysisReproducibility Assessment
PURPOSE: To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. METHODS: In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. RESULTS: Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. CONCLUSION: Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.
32
A H BECKETT, M ROWLAND, P TURNER · The Lancet · 1965 · 16.9K citations
Statistics Notes: Measurement error
J M. Bland, D. G Altman · BMJ · 1996 · 1.1K citations · Full text
Health Professions, Biostatistics, Individual Differences +16
Wolfgang Haigis, Barbara A.M. Lege, Naomi R. Miller et al. · Graefe s Archive for Clinical and Experimental Ophthalmology · 2000 · 765 citations
Sources of error in intraocular lens power calculation
Sverker Norrby · Journal of Cataract & Refractive Surgery · 2008 · 656 citations
Medicine, Ophthalmology, Surgery +17