Optimized cell geometry for buffer-gas-cooled molecular-beam sources

Vijay Pal Singh, Amit K. Samanta, Nils Roth, Daniel Gusa, Tim Ossenbrüggen, I. Rubinsky, Daniel A. Horke, Jochen Küpper

Physical review. A/Physical review, A · 2018 · 21 citations · 34 references

DOIFull text

Open access

Concepts

Abstract

We have designed, constructed, and commissioned a cryogenic helium buffer-gas source for producing a cryogenically cooled molecular beam and evaluated the effect of different cell geometries on the intensity of the produced molecular beam, using ammonia as a test molecule. Planar and conical entrance and exit geometries are tested. We observe a threefold enhancement in the ${\mathrm{NH}}_{3}$ signal for a cell with planar entrance and conical-exit geometry, compared to that for a typically used ``boxlike'' geometry with planar entrance and exit. These observations are rationalized by flow field simulations for the different buffer-gas cell geometries. The full thermalization of molecules with the helium buffer gas is confirmed through rotationally resolved resonance-enhanced multiphoton ionization spectra yielding a rotational temperature of 5 K.

References

34