Journal of Conservative Dentistry · 2017 · 13 citations · 21 references
Natural SciencesMaterials EngineeringMaterials ScienceBiomaterialsDental MaterialsProsthodontic MaterialsBarrier PropertiesMold ControlComposite Specimen EvaluationHardnessHardness Ratio AnalysisLight IntensityLight-curing EffectsLight-curing Intensity EffectsLight-curing SystemsSpecimen Measurement
AIMS: This study evaluated effect of infection control barriers on light intensity (LI) of light-curing unit (LCU) and microhardness of composite. MATERIALS AND METHODS: = 30) were tested against the control. LI for each barrier was measured with Lux meter. One hundred and fifty Teflon molds were equally divided into five groups of thirty each. Composite was filled in bulk in these molds and cured without and with barrier. Microhardness was evaluated on top and bottom surface of composite specimen with microhardness testing machine and hardness ratio (HR) was derived. STATISTICAL ANALYSIS USED: -test using SPSS version 18 software. RESULTS: < 0.05). However, HR determined inadequate curing only with LCGP. CONCLUSIONS: Although entire tested barrier significantly reduced the LI; none, except LCGP markedly affected the degree of cure of the composite.
21
Guidelines for Infection Control in Dental Health Care Settings
William Kohn, Amy S. Collins, Jennifer L. Cleveland et al. · Elsevier eBooks · 2013 · 698 citations