2017 International conference on Microelectronic Devices, Circuits and Systems (ICMDCS) · 2017 · 23 citations · 9 references
EngineeringBioengineeringBiomedical EngineeringAnalog ElectronicsBioinstrumentationNeural EngineeringAnalog Integrated CircuitsBiomedical InstrumentationCmos Circuit DesignLow-noise Circuit DesignNeural Recording TechnologiesNoise In Analog CircuitsBiomedical ApplicationsCmos Analog Integrated CircuitsIntegrated Circuit NoiseNeural RecordingNeural Recording SystemsNeural Signal AmplificationNoise In AmplifiersNoise In Integrated Circuits
In this paper, a PVT insensitive variable gain amplifier (VGA) has been proposed which is suitable for bio-medical neural recording. Proposed amplifier delivers an appreciably enhanced performance over that of existing technique. Constant Gm bias stability is described in the frequency domain. Noise problems with the existing technique are described qualitatively. A prototype circuit has been implemented in 65nm CMOS technology and simulated with Spectre. 2–25dB gain programmable range was achieved across PVT and 0.3dB standard deviation. 1.2uV integrated RMS noise was achieved. The circuit consumes 0.5mA from the 1.5V supply voltage and occupies 4500um <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> silicon area.
9