Publication | Closed Access
256-Channel Neural Recording and Delta Compression Microsystem With 3D Electrodes
184
Citations
21
References
2009
Year
Medical ElectronicsEngineeringNeural RecodingBiomedical EngineeringIntegrated CircuitsWireless Implantable DeviceNeurochipMedical InstrumentationSocial SciencesMum CmosBiomedical DevicesMicrofluidicsMm Platinum-coated MicroelectrodesBio-electronic InterfacesDelta Compression MicrosystemMicroelectronicsNeural InterfaceNeural InterfacesBiomedical SensorsNeuroengineeringCircuit DesignMicrofabricationBioelectronicsBiomedical InstrumentationElectrophysiologyNeuroscienceBrain Electrophysiology
A 3D microsystem for multi-site penetrating extracellular neural recording from the brain is presented. A 16 times 16-channel neural recording interface integrated prototype fabricated in 0.35 mum CMOS occupies 3.5 mm times 4.5 mm area. Each recording channel dissipates 15 muW of power with input-referred noise of 7 muV <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">rms</sub> over 5 kHz bandwidth. A switched-capacitor delta read-out data compression circuit trades recording accuracy for the output data rate. An array of 1.5 mm platinum-coated microelectrodes is bonded directly onto the die. Results of in vitro experimental recordings from intact mouse hippocampus validate the circuit design and the on-chip electrode bonding technology.