Journal of the American Society for Mass Spectrometry · 2017 · 66 citations · 31 references
PharmacyPharmaceutical AnalysisPharmaceuticsAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisChromatographyInstrumental AnalysisMass SpectrometryToxicological Drug ScreeningConcentrationIllicit Drug DetectionMass Spectrometric Drug AnalysisMolecularly Imprinted MaterialsMolecularly Imprinted PolymersPrecisionTandem Mass SpectrometryTandem Mass Spectrometry Analysis
This study proposes a new direct and fast method of analysis employing paper spray mass spectrometry (PS-MS). The paper used in the proposed method was modified with molecularly imprinted polymers (MIP) to create a specific site for cocaine analysis in oral fluid. MIP membrane was successfully synthetized and employed. The developed method showed to be linear in a concentration range from LOQ to 100 ng mL –1 . The experimental value of LOQ obtained was 1 ng mL –1 . The inter-day and intra-day precision and accuracy of the PS-MS method presented values lower than 15%. The total recoveries were also evaluated. The PS-MS method for the analysis of cocaine in oral fluid showed to be very promising and the validation parameters showed a good correlation with the literature.
31
Limit of detection. A closer look at the IUPAC definition
Gary L. Long, J. D. Winefordner · Analytical Chemistry · 1983 · 1.9K citations
Development of probe electrospray using a solid needle
Kenzo Hiraoka, Kentaro Nishidate, Kunihiko Mori et al. · Rapid Communications in Mass Spectrometry · 2007 · 328 citations